Abstract

Several problems such as metallic contamination due to amalgamation, capillary plugging and air leaks are associated with the use of the static hanging mercury drop electrode (SHMDE). This paper describes the design and construction of an improved version of this type of electrode. The modifications include an improved valve seat and capillary attaching assembly.